Due to the high gain and bandwidth of the device a single photon yields an output pulse with an amplitude in the range of 100 to 200 mV and a pulse width of 1.5 ns. The TCSPC instrument response function (IRF) has a width of less than 120 ps FWHM. It decreases to less than 100 ps for small sizes of the illuminated spot on the photocathode. The detector is operated with a fixed gain and from a single +12 V power supply. Overload conditions are internally detected by sensing the PMT output current. When overload occurs the PMT operating voltage is regulated down so that the output current remains on a safe level. Alternatively, the PMCS-150 can be operated from a bh DCU series detector controller module. In that case, the controller provides for power supply, gain control, and overload shutdown. The PMCS-150 interfaces directly to all bh SPC series time-correlated single photon (TCSPC) devices.