The PMC-150 contains a fast miniature PMT along with a Peltier cooler, a high voltage generator, a GHz pulse amplifier and a current sensing circuit. Due to the high gain and bandwidth of the device a single photon yields an output pulse with an amplitude in the range of 100 to 200 mV and a pulse width of 1.5 ns. Due to the high gain and the efficient shielding noise pickup is minimised. Therefore the PMC-150 yields high time resolution and high counting efficiency. The TCSPC instrument response function (IRF) has a width of less than 150 ps FWHM. Overload conditions are detected by sensing the PMT output current. Overload is indicated by an LED, an acaustic signal, and a logical overload signal. The PMC-150 is operated by the bh DCC-100 detector controller card. The DCC delivers the operting voltage for the PMT, the current for the Peltier cooler, controls the detector gain, and shuts down the PMT in case of overload. Compared to its predecessor, the PMC-100, the PMC-150 has a shorter IRF width and a better IRF uniformity over the active
area.
Main Features:
- Fast PMT for Single Photon Counting Applications
- Fast TCSPC Instrument Response: Typ. 130 ps FWHM
- Internal Cooler: Low Dark Count Rate
- Internal GHz Preamplifier: High Output Amplitude
- Fast Overload Protection
- Higher Gain Compared to its Predecessor PMC-100
- Excellent Noise Immunity
- Standard C-Mount Thread Connection
- Integrated High-Voltage Generator
- Operated via DCC-100 or DCU-400/DCU-800 Laser / Detector Controller
- Direct Interfacing to All bh TCSPC Systems